Skip to main content
Dryad

Application of ATLD-MCR to gas chromatography-mass spectrometric data for the quantification of PAHs in aerosols

Data files

Feb 21, 2022 version files 28.87 MB

Click names to download individual files

Abstract

In the work, for the first time, alternating trilinear decomposition-assisted multivariate curve resolution (ATLD-MCR) was applied to analyse complex gas chromatography–mass spectrometric (GC-MS) data with severe baseline drifts, serious co-elution peaks and slight retention time shifts for the simultaneous identification and quantification of polycyclic aromatic hydrocarbons (PAHs) in aerosols. It was also compared with the classic multivariate curve resolution-alternating least-squares (MCR-ALS) and the GC-MS-based external standard method. In validation samples, average recoveries of five PAHs were within the range from (96.2 ± 6.8)% to (106.5 ± 4.1)% for ATLD-MCR, near to the results of MCR-ALS ((98.0 ± 1.5)% to (106.7 ± 4.3)%). In aerosol samples, the concentrations of pyrene provided by ATLD-MCR were not significantly different from those of MCR-ALS. The other four PAHs including chrysene, benzo[a]anthracene, fluoranthene and benzo[b]fluoranthene were not detected by ATLD-MCR and the GC-MS-based external standard method. The results of figures of merit further demonstrated that ATLD-MCR achieved high sensitivities (8.9 × 104 to 1.7 × 106 mAU ml µg−1) and low limits of detection (0.003 to 0.087 µg ml−1), which were better than or similar to MCR-ALS, presenting a great choice to deal with complex GC-MS data for the simultaneous determination of targeted PAHs in aerosols.