Data from: Investigation of parallel radiofrequency transmission for the reduction of heating in long conductive leads in 3 tesla magnetic resonance imaging

McElcheran CE, Yang B, Golenstani-Rad L, Graham SJ

Date Published: August 5, 2015

DOI: http://dx.doi.org/10.5061/dryad.53bh1

 

Files in this package

Content in the Dryad Digital Repository is offered "as is." By downloading files, you agree to the Dryad Terms of Service. To the extent possible under law, the authors have waived all copyright and related or neighboring rights to this data. CC0 (opens a new window) Open Data (opens a new window)

Title Axial Image with 4-element PTX coil in suppression mode
Downloaded 6 times
Description Axial slice through center of phantom obtained with GRE sequence (TR = 2000 ms, TE = 40 ms, slice thickness = 10 mm, in-plane resolution = 256x256, FOV = 24 cm) with 4-Element PTX operating in suppression mode (phase shifts of 40, 16, 200 and 301 degrees). Dicom file for image shown in manuscript.
Download AxialGREwPTX.dcm (145.2 Kb)
Details View File Details
Title Coil Geometry
Downloaded 6 times
Description Contains specs about the geometry of the coil configurations (including capacitance values) for the simulations.
Download CoilGeometry.txt (466 bytes)
Details View File Details
Title Birdcage Heating
Downloaded 6 times
Description Temperature probe measurements for heating at tip of wire 12 cm long copper wire in a polyacrylic acid cylindrical phantom (radius = 9 cm, height = 24 cm) using FRFSE sequence (TR = 559 ms, TE = 100 ms, echo train length = 24, 1 slice, imaging time = 4:08). Excitation with transmit/receive GE birdcage coil with transmit gain of 153.
Download Birdcage Heating Test 1 TX153.txt (176.2 Kb)
Details View File Details
Title PTx Quadrature Heating
Downloaded 3 times
Description Temperature probe measurements for heating at tip of wire 12 cm long copper wire in a polyacrylic acid cylindrical phantom (radius = 9 cm, height = 24 cm) using FRFSE sequence (TR = 559 ms, TE = 100 ms, echo train length = 24, 1 slice, imaging time = 4:08). Excitation with 4-element parallel transmit coil with phase shifts of 0, 90, 180 and 270 degrees between elements with transmit gain of 195
Download PTx Quad Heating Test 1 TX195.txt (143.3 Kb)
Details View File Details
Title PTx Suppression Heating Test 1
Downloaded 4 times
Description Temperature probe measurements for heating at tip of wire 12 cm long copper wire in a polyacrylic acid cylindrical phantom (radius = 9 cm, height = 24 cm) using FRFSE sequence (TR = 559 ms, TE = 100 ms, echo train length = 24, 1 slice, imaging time = 4:08). Excitation with 4-element parallel transmit coil with phase shifts of 40, 16, 200 and 301 degrees (suppression mode) with a transmit gain of 190
Download PTx Suppression Heating Test 1 TX190.txt (149.5 Kb)
Details View File Details
Title PTx Suppression Heating Test 2
Downloaded 5 times
Description Temperature probe measurements for heating at tip of wire 12 cm long copper wire in a polyacrylic acid cylindrical phantom (radius = 9 cm, height = 24 cm) using FRFSE sequence (TR = 559 ms, TE = 100 ms, echo train length = 24, 1 slice, imaging time = 4:08). Excitation with 4-element parallel transmit coil with phase shifts of 40, 16, 200 and 301 degrees (suppression mode) with a transmit gain of 187. Second measurement.
Download PTx Suppression Heating Test 2 TX187.txt (144.7 Kb)
Details View File Details
Title E-field increase as function of wire position
Downloaded 35 times
Description Folder containing two files with % increase in E-field magnitude at tip of wire vs minimum in field of view for varying radial position (RposRatio.csv) and z position (ZposRatio.csv)
Download WirePos.zip (664 bytes)
Details View File Details
Title Simulation data for wire at r = 2.5 cm, z = 0 cm (E-field and H-field)
Downloaded 31 times
Description Contains simulation data for wire positions r = 2.5 cm, z = 0 cm. Includes data for a 16-rung birdcage coil operating in quadrature mode and a 2-element PTX coil, 4-element PTX coil and 8-element PTX operating in suppression mode (amplitudes and phases found in manuscript). Each folder contains a folder named 'XY plane' corresponding to data in the z = 0 cm plane, x = -10 cm to x = 10 cm, y = -10 cm to 10 cm and 'YZ plane' corresponding to data in the x = 0 cm plane, z = -12 cm to 1 cm, y = -10 cm to 10 cm. Each file is named: <Coil config><field><plane>_<radial pos of wire>.txt. The files are space delimited. Header of 15 lines.
Download Leadat25mm.zip (7.026 Mb)
Details View File Details
Title Simulation data for wire at r = 8 cm, z = 2 cm (E-field and H-field)
Downloaded 35 times
Description Contains simulation data for wire positions r = 8 cm, z = 2 cm. Includes data for a 16-rung birdcage coil operating in quadrature mode and a 2-element PTX coil, 4-element PTX coil and 8-element PTX operating in suppression mode (amplitudes and phases found in manuscript). Each folder contains a folder named 'XY plane' corresponding to data in the z = 0 cm plane, x = -10 cm to x = 10 cm, y = -10 cm to 10 cm and 'YZ plane' corresponding to data in the x = 0 cm plane, z = -12 cm to 1 cm, y = -10 cm to 10 cm. Each file is named: <Coil config><field><plane>_<radial pos of wire>.txt. The files are space delimited. Header of 15 lines
Download Leadat80mmz2cm.zip (12.87 Mb)
Details View File Details
Title Simulation data for wire at r = 7.5 cm, z = 0 cm (E-field and H-field)
Downloaded 30 times
Description Contains simulation data for wire positions r = 7.5 cm, z = 0 cm. Includes data for a 16-rung birdcage coil operating in quadrature mode and a 2-element PTX coil, 4-element PTX coil and 8-element PTX operating in suppression mode (amplitudes and phases found in manuscript). Each folder contains a folder named 'XY plane' corresponding to data in the z = 0 cm plane, x = -10 cm to x = 10 cm, y = -10 cm to 10 cm and 'YZ plane' corresponding to data in the x = 0 cm plane, z = -12 cm to 1 cm, y = -10 cm to 10 cm. Each file is named: <Coil config><field><plane>_<radial pos of wire>.txt. The files are space delimited. Header of 15 lines
Download Leatat75mm.zip (7.003 Mb)
Details View File Details

When using this data, please cite the original publication:

McElcheran CE, Yang B, Golenstani-Rad L, Graham SJ (2015) Investigation of parallel radiofrequency transmission for the reduction of heating in long conductive leads in 3 tesla magnetic resonance imaging. PLOS ONE 10(8): e0134379. http://dx.doi.org/10.1371/journal.pone.0134379

Additionally, please cite the Dryad data package:

McElcheran CE, Yang B, Golenstani-Rad L, Graham SJ (2015) Data from: Investigation of parallel radiofrequency transmission for the reduction of heating in long conductive leads in 3 tesla magnetic resonance imaging. Dryad Digital Repository. http://dx.doi.org/10.5061/dryad.53bh1