Skip to main content
Dryad

Data from: Optimal control pulses for the 1.2 GHz (28.2 T) NMR spectrometers

Data files

Oct 16, 2023 version files 238.42 KB

Click names to download individual files

Abstract

The ability to measure NMR spectra with a large sample volume is crucial for concentration limited biological samples in order to attain adequate signal-to-noise (S/N) ratio. The possibility to measure with a 5 mm cryoprobe is currently absent at the 1.2 GHz NMR instruments due to the exceedingly high radio frequency power demands, which is four times compared to a 600 MHz instruments. Here, we overcome the high power demands by designing optimal control (OC) pulses with up to 20 times lower power requirements than currently necessary at a 1.2 GHz spectrometer. We show that multidimensional biomolecular NMR experiments constructed using these OC pulses can bestow improvement in the S/N ratio of up to 26%. With the expected power limitations of a 5 mm cryoprobe, we observe an enhancement in the S/N ratio of more than 240% using our OC sequences. This motivates the development of a cryoprobe with a larger volume than the current 3 mm cryoprobes.