Ultrasound neural interface enabling acute spinal cord monitoring and blood flow modulation
Data files
Oct 01, 2026 version files 8.93 MB
-
MUSIC_FUS_data.zip
8.91 MB
-
README.md
20.20 KB
Abstract
Dryad DOI: https://doi.org/10.5061/dryad.nk98sf891
Associated manuscript: Ultrasound neural interface enabling acute spinal cord monitoring and blood flow modulation
Journal: Science Advances
Manuscript ID: aeg2342
Dataset summary
This Dryad dataset contains the experimental spectral-Doppler time-series data used for the focused-ultrasound (FUS) blood-flow-velocity (BFV) analyses in the associated study of the Multimodal Ultrasound System for Injury Care (MUSIC).
The study evaluated whether MUSIC-delivered focused ultrasound could modulate Doppler-derived spinal cord blood-flow velocity in healthy porcine experiments. Across three porcine experiments, FUS parameters were varied, including input voltage, pulse repetition frequency (PRF), pulse/burst duration, and insonation duration. The analyses compared FUS-off spectral-Doppler signals immediately before and after FUS delivery. FUS-on intervals were generally excluded from the main pre/post analysis because the FUS pulses could generate spectral interference.
The uploaded archive is:
MUSIC_FUS_data.zip
It contains two organizational views of the same underlying data:
Original/: the canonical source organization, preserving the chronological experimental sections.Further_class/: byte-identical copies of theOriginal/MATLAB files reorganized by experimental parameter grouping for convenient downstream analysis.
Important: Further_class/ does not contain additional independent observations. It is a reorganized copy of the source data in Original/. The archive contains 38 source .mat files in Original/ and 78 reorganized copies in Further_class/. Hash comparison confirms that every .mat file in Further_class/ is identical to one of the 38 source files in Original/.
The 38 source files define 30 repeated FUS sessions across three porcine experiments. A FUS session is represented by two adjacent FUS-off recordings: the end of one recording is used as the pre-FUS endpoint and the beginning of the next recording as the post-FUS endpoint.
Relationship to the manuscript
These data support the proof-of-concept FUS/BFV analyses described in the main manuscript and Supplementary Materials, including the parameter-sweep and pre/post-FUS analyses associated with main Figure 5 and Supplementary Figures S9-S12.
The associated study interprets these measurements as Doppler-derived local hemodynamic measurements. They should not be interpreted as direct measurements of vessel diameter, volumetric blood flow, vasodilation, tissue perfusion, or therapeutic efficacy.
Description of the data and file structure
Top-level structure
MUSIC_FUS_data/
├── Original/
│ ├── 0307/
│ │ └── Replicate/
│ ├── 0627/
│ │ ├── a.5KHz Control Section/
│ │ ├── b. Duty Cycle Control Section/
│ │ └── c. Long FUS Section/
│ ├── 0729/
│ │ ├── a. MAX Section/
│ │ ├── b. Replicate Section/
│ │ └── c. Long FUS Section/
│ └── FUS6.mat
└── Further_class/
├── 3.7/
├── 6.27/
├── 6.27duty/
├── 6.27voltage/
├── 7.29/
└── 7.29duration/
The labels 0307, 0627, and 0729 identify the three porcine experimental sessions in the archived analysis workflow. The corresponding parameter-organized labels are 3.7, 6.27, and 7.29.
Original/: canonical chronological source data
The Original/ directory contains the 38 source MATLAB files used to construct the analysis dataset.
| Folder | Number of .mat files |
Number of adjacent pre/post FUS sessions | Experimental context |
|---|---|---|---|
0307/Replicate/ |
5 | 4 | Replicate FUS trials from the first porcine experiment. Parameter-organized copies divide these into 1-kHz and 5-kHz groups at 60 V and a .07ms setting. |
0627/a.5KHz Control Section/ |
7 | 6 | Input-voltage/pressure-dose section from the second porcine experiment. Parameter-organized copies group adjacent trials at 20 V, 30 V, and 45 V with a 5-kHz and .07ms setting. |
0627/b. Duty Cycle Control Section/ |
7 | 6 | Pulse/burst-duration or duty-dose section from the second porcine experiment. Parameter-organized copies group adjacent trials at .15ms, .2ms, and .4ms settings with 1 kHz and 40 V. |
0627/c. Long FUS Section/ |
2 | 1 | Single longer-duration FUS comparison from the second porcine experiment. |
0729/a. MAX Section/ |
7 | 6 | Experimenter-labeled MAX section from the third porcine experiment. Parameter-organized copies include 1kHz 20V .35ms and 5kHz 20V .4ms groups and selected 2-minute comparisons. |
0729/b. Replicate Section/ |
7 | 6 | Replicate FUS trials from the third porcine experiment. Parameter-organized copies divide the sequence into 1-kHz and 5-kHz groups at 20 V and a .07ms setting. |
0729/c. Long FUS Section/ |
2 | 1 | Long-duration FUS comparison from the third porcine experiment; the reorganized duration folder labels this comparison 10min. |
Original/FUS6.mat |
1 | — | Source recording used as the final boundary recording for the 0627 voltage-series reorganization. It is paired with the preceding recording when forming the final voltage-comparison session. |
Total: 38 source MATLAB files defining 30 adjacent pre/post FUS sessions.
How adjacent files define a FUS session
Within each chronological section, files are ordered baseline/recovery recordings around FUS applications.
For a sequence such as:
FUSoff0.mat
FUSoff1.mat
FUSoff2.mat
the analysis treats:
- the end of
FUSoff0.matas the pre-FUS endpoint for session 1; - the beginning of
FUSoff1.matas the post-FUS endpoint for session 1; - the end of
FUSoff1.matas the pre-FUS endpoint for session 2; and - the beginning of
FUSoff2.matas the post-FUS endpoint for session 2.
Thus, n + 1 ordered FUS-off recordings define n adjacent FUS sessions.
Some legacy source files in Original/ are named FUS0.mat, FUS1.mat, etc. Their parameter-organized copies in Further_class/ are used as FUS-off boundary recordings in the analysis. Users should therefore rely on the chronological ordering and the parameter-organized grouping rather than interpret the legacy FUS prefix as indicating that the archived time series itself represents a FUS-on interval.
Further_class/: parameter-organized copies
Further_class/ contains no new measurements. It reorganizes copies of the source .mat files so that adjacent recordings corresponding to a particular parameter comparison are located together.
Further_class/3.7/
3.7 1kHz 60V .07ms/: three ordered recordings defining two adjacent sessions.3.7 5kHz 60V .07ms/: three ordered recordings defining two adjacent sessions.
These files are copies of Original/0307/Replicate/.
Further_class/6.27/
This directory provides grouped parameter sequences from the second porcine experiment:
6.27 5kHz 20V .07ms/6.27 5kHz 30V .07ms/6.27 5kHz 45V .07ms/6.27 1kHz 40V .15ms/6.27 1kHz 40V .2ms/6.27 1kHz 40V .4ms/6.27 Long FUS Section/
The numeric strings in these folder names are experimental-setting labels:
kHz: pulse repetition frequency (PRF) label, in kilohertz;V: FUS system input-voltage setting, in volts;ms: pulse/burst-duration setting, in milliseconds.
Further_class/6.27voltage/
This folder expands the voltage-series data into individual adjacent pre/post pairs:
- 20-V trials: folders labeled
20V .07msand20V .07ms_1; - 30-V trials: folders labeled
30V .07msand30V .07ms_1; - 45-V trials: folders labeled
45V .07msand45V .07ms_1.
Each two-file folder defines one adjacent FUS session. The suffixes _1 and _2 are replicate/trial identifiers and do not represent physical units.
Further_class/6.27duty/
This folder expands the pulse/burst-duration series into individual adjacent pre/post pairs:
.15msand.15ms_2;.2msand.2ms_2;.4msand.4ms_2.
Each two-file folder defines one adjacent FUS session. These comparisons were used to investigate dose dependence related to pulse/burst-duration and duty-cycle settings.
Further_class/7.29/
This directory provides parameter-organized copies from the third porcine experiment:
7.29 1kHz 20V .07ms/7.29 5kHz 20V .07ms/7.29 Max 1kHz 20V .35ms/7.29 Max 5kHz 20V .4ms/7.29 Short FUS 2min/7.29 Short FUS 2min_2/7.29 Long FUS Section/
MAX is retained as the experimenter-defined section label. The actual numeric experimental-setting labels are preserved in the folder names.
Further_class/7.29duration/
This folder isolates the duration comparison:
7.29 Short FUS 2min_1/: two recordings defining a 2-minute comparison session.7.29 zLong FUS 10min/: two recordings defining a 10-minute comparison session.
Two additional .mat files at the root of 7.29duration/ are duplicate copies of the same source recordings used in the short-duration comparison and do not constitute additional sessions.
MATLAB file contents and variable dictionary
Each .mat file contains one primary one-dimensional spectral-Doppler envelope time series plus an extraction-threshold scalar and MATLAB object metadata.
Two naming conventions occur because the source data were exported in different experimental sessions. US_chunk and US1 represent the same logical type of signal; similarly, thresh and thresh1 represent the same logical type of saved extraction parameter.
| Variable | Type / shape | Units or coding | Meaning |
|---|---|---|---|
US_chunk |
1 × N numeric array (float64) |
Hz | Extracted spectral-Doppler envelope/frequency-shift time series. This is the primary analysis signal in files using the US_chunk naming convention. |
US1 |
1 × N numeric array (float64) |
Hz | Same logical signal type as US_chunk, used in a subset of source files created under a different export naming convention. |
thresh |
1 × 1 numeric scalar (float64) |
Unitless extraction parameter | Threshold value saved with files using US_chunk. In this archive values range from 0.05 to 0.775. It is retained for provenance; the archived Python pre/post analysis does not use this scalar as an analysis outcome. |
thresh1 |
1 × 1 numeric scalar (float64) |
Unitless extraction parameter | Same logical type as thresh, paired with files using US1. |
None |
MATLAB object metadata | Not an analysis variable | Serialization metadata associated with the MATLAB datetime object saved during export (t_US_chunk). It is not used in the archived Python BFV analysis and should not be interpreted as a physiological variable. |
Across the canonical Original/ source files:
- 26 files store the primary signal as
US_chunk; - 12 files store the primary signal as
US1.
The length of the primary signal varies by recording because recording duration varied across experimental sections. In the canonical source files, array lengths range from 5,886 to 25,836 samples.
Sampling and time windows
The archived analysis workflow uses 1,000 samples for the 20-second pre-FUS and post-FUS comparison windows, corresponding to an exported envelope sampling rate of approximately 50 samples/s (50 Hz).
For each FUS session:
- pre-FUS window: the final 1,000 samples (20 s) of the preceding FUS-off recording;
- post-FUS window: the first 1,000 samples (20 s) of the following FUS-off recording.
For selected supplementary analyses, expanded approximately 30-second FUS-off windows were used as described in the manuscript.
Interpretation of the primary signal
US_chunk/US1 contains the extracted spectral-Doppler envelope expressed as Doppler frequency shift in Hz. The manuscript uses these spectral-Doppler data to assess relative and normalized changes in Doppler-derived blood-flow velocity across pre- and post-FUS conditions.
The archived frequency-shift series should not be interpreted as a direct measurement of volumetric blood flow or tissue perfusion.
File-name and folder-name interpretation
| Label/pattern | Interpretation |
|---|---|
0307, 0627, 0729 |
Internal experimental-session/date labels corresponding to the three porcine experiments. |
3.7, 6.27, 7.29 |
Parameter-organized versions of the same three experimental-session labels. |
FUSoffN, offN |
Ordered FUS-off baseline/recovery recording. N indicates position in the experimental sequence. |
FUSN in selected Original/ folders |
Legacy source filename. In the reorganized analysis structure, these files are used as FUS-off boundary recordings. |
1kHz, 5kHz |
PRF setting label in kilohertz. |
20V, 30V, 40V, 45V, 60V |
FUS system input-voltage setting in volts. |
.07ms, .15ms, .2ms, .35ms, .4ms |
Pulse/burst-duration setting label in milliseconds. |
2min, 10min |
FUS insonation-duration label in minutes. |
_1, _2 |
Replicate/trial grouping suffix; no physical unit. |
Replicate |
Experimental section containing repeated FUS trials. |
Duty Cycle Control Section |
Experimental section used to compare pulse/burst-duration and related duty-dose settings. |
Long FUS Section |
Experimental section used for longer-duration FUS comparison. |
MAX Section / Max |
Experimenter-defined section label retained from the source dataset; numeric settings are provided separately in parameter-organized folder names. |
Missing/invalid values and analysis handling
The .mat files do not use a dedicated textual missing-value code.
In the archived Python analysis workflow:
- zero-valued points in the extracted envelope can be treated as invalid/missing envelope samples;
- invalid points are converted to missing values and interpolated;
- extreme outliers and abrupt signal jumps may be removed/interpolated as part of preprocessing;
- smoothing and pre/post window extraction are then applied for the relevant analyses.
Because exact preprocessing differs by analysis/figure, users seeking to reproduce the manuscript should use the archived publication-version software listed below rather than reimplement preprocessing solely from this README.
Experimental context and analysis logic
The focused-ultrasound experiments were conducted in three healthy porcine experiments. During each experimental section, FUS parameters were varied to evaluate whether dose-related settings were associated with changes in local Doppler-derived BFV.
The principal experimental variables included:
- FUS system input voltage;
- pulse repetition frequency (PRF);
- pulse/burst-duration or duty-related settings; and
- insonation duration.
A typical multi-session section consisted of repeated approximately 2-minute FUS sessions separated by FUS-off baseline/recovery recordings. Selected sections included a longer approximately 10-minute FUS exposure. Recovery time was provided between experimental sections.
The main pre/post analysis uses only FUS-off windows immediately before and after therapy. This avoids the harmonic interference that can be present in spectral-Doppler recordings during FUS delivery.
For a session-level percentage-change calculation, the archived software computes:
100 × (mean_post - mean_pre) / mean_pre
where mean_pre and mean_post are calculated from the selected pre- and post-FUS windows.
These repeated sessions are within-animal observations and should not be treated as 30 independent animals.
Sharing/access information
Permanent experimental-data archive
Dryad:
https://doi.org/10.5061/dryad.nk98sf891
Permanent publication-version software archives
MUSIC FUS Analysis
Zenodo: https://doi.org/10.5281/zenodo.22907737
GitHub development repository: https://github.com/HEPIUSLAB/MUSIC_FUS_Analysis
HEPIUS MUSIC Imaging
Zenodo: https://doi.org/10.5281/zenodo.22908234
GitHub development repository: https://github.com/HEPIUSLAB/HEPIUS_MUSIC_imaging
MUSIC Mobile App
Zenodo: https://doi.org/10.5281/zenodo.22908477
GitHub development repository: https://github.com/HEPIUSLAB/MUSIC_APP
HEPIUS Lab GitHub organization:
https://github.com/HEPIUSLAB
Code/software
The software used to organize, preprocess, analyze, and visualize the FUS spectral-Doppler data is archived permanently in the MUSIC FUS Analysis Zenodo record:
https://doi.org/10.5281/zenodo.22907737
The archived software includes Python source code and Jupyter notebooks. The analysis code recognizes either US_chunk or US1 as the primary spectral-Doppler envelope variable and implements the pre/post pairing logic described above.
The GitHub repository is maintained as a development resource; the Zenodo DOI should be used when citing the publication-version software.
Licensing
Data deposited in Dryad are released under Dryad's applicable CC0 waiver.
Software is licensed separately according to the license stated in each Zenodo software record.
Abbreviations
- BFV: blood-flow velocity
- FUS: focused ultrasound
- MUSIC: Multimodal Ultrasound System for Injury Care
- PRF: pulse repetition frequency
- RI: resistivity index
- SCI: spinal cord injury